Feasibility and analysis of CNN-based candidate beam generation for robotic radiosurgery.

Abstract

The coverage increases from 95.35% to 97.67% for 6000 heuristically and CNN-generated candidate beams, respectively. Conversely, a similar coverage can be achieved for treatment plans with half the number of candidate beams. This results in a patient-dependent reduced averaged computation time of 20.28%-45.69%. The number of active treatment beams can be reduced by 11.35% on average, which reduces treatment time. Constraining the maximum number of candidate beams per beam node can further improve the average coverage by 0.75 percentage points for 6000 candidate beams.

We show that deep learning based on radiological features can substantially improve treatment plan quality, reduce computation runtime, and treatment time compared to the heuristic approach used in clinics.

In clinical practice, a set of candidate beams generated by a randomized heuristic forms the basis for treatment planning. We use a convolutional neural network to identify promising candidate beams. Using radiological features of the patient, we predict the influence of a candidate beam on the delivered dose individually and let this prediction guide the selection of candidate beams. Features are represented as projections of the organ structures which are relevant during planning. Solutions to the inverse planning problem are generated for random and CNN-predicted candidate beams.

© 2020 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.